CALibrate:DPRog (@<channel>)
Calibrates the current downprogrammer.
Parameter Typical Return
(none) (none)
Calibrates the current downprogrammer: CAL:DPR (@1)
CALibrate:LEVel P1|P2|P3
Advances to the next level in the calibration. P1 is the first level; P2 is the second, P3 is the third.
Parameter Typical Return
P1|P2|P3 (none)
Selects the first calibration level: CAL:LEV P1
l Some calibration sequences may require some settling time after sending CAL:LEV but before
reading the data from the external meter and sending CAL:DATA.
CALibrate:PASSword <password>
Sets a numeric password to prevent unauthorized calibration. The new password is automatically
stored in nonvolatile memory and does not have to be stored with CALibrate:SAVE.
Parameter Typical Return
<password>a numeric value up to 15 digits (none)
Set a new password to a value of 1234: CAL:PASS1234
l If the password is set to 0, password protection is removed and the ability to enter calibration
mode is unrestricted. The factory setting is 0 (zero).
l To change the password: unsecure calibration with the old password, then set the new password.
l If you forget your password, refer to the calibration section.
CALibrate:RESistance 20|6, (@<channel>)
Calibrates the output resistance. The parameter selects the voltage range. Note that the 6 V range
only applies to Model N6781A.
Parameter Typical Return
20|6 (none)
Calibrates the outputresistanceof the 20 V range: CAL:RES 20, (@1)
Keysight N6705C Operating and Service Guide 269