[SOURce:]CURRent:PROTection:DELay:STARt SCHange|CCTRans, (@<chanlist>)
[SOURce:]CURRent:PROTection:DELay:STARt? (@<chanlist>)
Specifies what starts the over-current protection delay timer.
SCHange starts the over-current delay timer at the END of a settings change in voltage, current, or
output state; allowing for additional protection delay time. Transitions into constant current mode are
ignored DURING a programmed settings change in voltage, current, or output state.
CCTRans starts the over-current delay timer by any transition of the output into current limit mode.
Parameter Typical Return
SCHange|CCTRans, *RST SCHange SCH or CCTR
Selects the CCTRans as the current protection delay mode:
CURR:PROT:DEL:STAR CCTR, (@1)
[SOURce:]CURRent:PROTection:STATe 0|OFF|1|ON, (@<chanlist>)
[SOURce:]CURRent:PROTection:STATe? (@<chanlist>)
Enables or disables the over-current protection. If the over-current protection function is enabled and
the output goes into constant current operation, the output is disabled and the Questionable
Condition status register OCP bit is set.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enable the current protection state: CURR:PROT:STAT ON, (@1)
l The current limit setting determines when the output goes into constant current operation.
l An over-current condition can be cleared with OUTPut:PROTection:CLEar after the cause of the
condition is removed.
[SOURce:]CURRent:RANGe <value>, (@<chanlist>)
[SOURce:]CURRent:RANGe? (@<chanlist>)
Sets the output current range on models that have multiple ranges. The value that you enter must be
the highest value in amperes that you expect to source. The instrument selects the range with the
best resolution for the value entered.
Parameter Typical Return
0 - maximum|MIN|MAX, *RST MAX
The maximum value is 102% of the current rating of
the power module.
<currentrange>
Sets the outputcurrent range to 3 A: CURR:RANG 3, (@1)
7 SCPI Programming Reference
274 Keysight N6705C Operating and Service Guide