Parameter Typical Return
RST|RCL0 RST or RCL0
Sets the power-on state to the *RST state: OUTP:PON:STAT RST
OUTPut:PROTection:CLEar (@<chanlist>)
This clears the latched protection status that disables the output when a protection condition occurs
(see Programming Output Protection). The output is restored to the state it was in before the
protection condition occurred.
Parameter Typical Return
(none) (none)
Clears the latched protection status: OUTP:PROT:CLE (@1)
l All conditions that generate the fault must be removed before the latched status can be cleared.
The output is restored to the state it was in before the fault condition occurred.
l If a protection shutdown occurs during an output list, the list continues running even though the
output is disabled. When the protection status is cleared and the output becomes enabled again,
the output will be set to the values of the step that the list is presently at.
OUTPut:PROTection:COUPle 0|OFF|1|ON
Enables or disables channel coupling for protection faults. When enabled, ALL output channels are
disabled when a protection fault occurs on any output channel. When disabled, only the affected
output channel is disabled when a protection fault is triggered
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0๎or๎1
Turns the outputcoupling protection on: OUTP:PROT:COUP๎ON
OUTPut:PROTection:DELay <value>, (@<chanlist>)
OUTPut:PROTection:DELay? [MIN|MAX,] (@<chanlist>)
This command is the same as [SOURce:]CURRent:PROTection:DELay.
Sets the over-current protection delay. The over-current protection function will not be triggered on
the specified output channel during the delay time. After the delay time has expired, the over-current
protection function will be active. This prevents momentary changes in output status from triggering
the over-current protection function. Programmed values can range from 0 to 255 milliseconds in
increments of 1 millisecond.
Keysight N6705C Operating and Service Guide 311
7 SCPI Programming Reference