SENSe:HISTogram:CURRent[:DC]:RANGe[:UPPer] <value>, (@<chanlist>)
SENSe:HISTogram:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects the current histogram measurement range. The value that you enter must be the highest
value that you expect to measure. Units are in amperes. The instrument selects the range with the
best resolution for the value entered.
Parameter Typical Return
0 to maximum|MIN|MAX, *RST MAX <maximum value of range>
Selects the 1 A range: SENS:HIST:CURR:RANG 1, (@1)
l When queried, the returned value is the maximum current that can be measured on the range that
is presently set.
SENSe:HISTogram:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON, (@<chanlist>)
SENSe:HISTogram:CURRent[:DC]:RANGe:AUTO? (@<chanlist>)
Enables or disables seamless histogram current measurement autoranging.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enable seamless histogram measurement autoranging:
SENS:HIST:CURR:RANG:AUTO ON, (@1)
l Autoranging does not include the 10 μA range, which must be selected individually.
SENSe:HISTogram:FUNCtion:CURRent <Bool>, (@<chanlist>)
SENSe:HISTogram:FUNCtion:CURRent? (@<chanlist>)
Enables or disables the current histogram function.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enables currenthistogram measurements: SENS:HIST:FUNC:CURR ON, (@1)
SENSe:SWEep:OFFSet:POINts <value>, (@<chanlist>)
SENSe:SWEep:OFFSet:POINts? [MIN|MAX,] (@<chanlist>)
Defines the offset in a data sweep for triggered measurements. Positive values represent the delay
after the trigger occurs but before the samples are acquired. Negative values represent data samples
taken prior to the trigger.
7 SCPI Programming Reference
324 Keysight N6705C Operating and Service Guide