SYSTem:PASSword:FPANel:RESet
Resets the front panel lockout password to zero. This command does not reset the calibration
password.
The front panel password can also be reset to 0 by setting an internal switch on the unit. Refer to
Calibration Switches for more information.
Parameter Typical Return
(none) (none)
To resetthe front panel password: SYST:PASS:FPAN:RES
SYSTem:PERSona:MANufacturer "Agilent"|"Keysight"
SYSTem:PERSona:MANufacturer?
SYSTem:PERSona:MANufacturer:DEFault
Changes the manufacturer identity. This command is used to change the manufacturer's ID string for
backward compatibility. This affects the following identity items:
l *IDN? contains "Agilent" instead of "Keysight"
l Default LAN hostname has the prefix A instead of K
l The LXI XML file fields include "Agilent" instead of "Keysight"
l The mDNS service name contains "Agilent" instead of "Keysight"
The DEFault command returns the manufacturer's ID back to Keysight.
Parameter Typical Return
"Agilent"|"Keysight" Keysight
Changes the manufacturer's ID to Agilent: SYST:PERS:MAN "Agilent"
SYSTem:PERSona:MODel "<model number>"
SYSTem:PERSona:MODel?
SYSTem:PERSona:MODel:DEFault
Changes the model number. This command is used to change the mainframe model ID string for
backward compatibility.
Compatibility Factory setting
N6705A|N6705B N6705C
The DEFault command returns the model's ID to the factory setting.
7 SCPI Programming Reference
336 Keysight N6705C Operating and Service Guide

Table of Contents