Parameter Typical Return
"<model number>" N6705C
Changes the model number ID to N6705B: SYST:PERS:MOD "N6705B"
SYSTem:REBoot
Reboots the instrument to its power-on state. Refer to Reset State for more information.
Parameter Typical Return
(none) (none)
To rebootthe instrument: SYST:REB
SYSTem:SECurity:IMMediate
Clears all user memory and reboots the instrument. This command is typically used to prepare the
instrument for removal from a secure area. It sanitizes all user data by writing all zeros to flash
memory and then performing a chip erase as per manufacturer's data sheet. Identification data
(instrument firmware, model number, serial number, MAC address and calibration data) is not erased.
After the data is cleared, the instrument is rebooted.
This procedure is not recommended for use in routine applications because of the possibility of
unintended loss of data.
Parameter Typical Return
(none) (none)
Sanitizes the instrument: SYST:SEC:IMM
SYSTem:SET <data>
SYSTem:SET?
Sets the instrument to the specified state. Used to get an instrument state and later restore the
instrument back to that state. <data> is a definite length arbitrary binary block of the form
#<D><DD...D><data-bytes>, where
#<D> indicates the response is definite length with <D> indicating the number of digits in the length
field.
<DD...D> indicates the length of the binary block containing the actual file.
<data-bytes> are the contents of the file.
The query returns a definite length binary block that contains the present instrument state.
Keysight N6705C Operating and Service Guide 337
7 SCPI Programming Reference