Front Panel Menu Reference SCPI Command
From the Calibration menu, select Miscellaneous, Downprogrammer. Select
Start. Then select Close.
CAL:DPRog (@1)
Enter a Calibration Date, Save, and Log Out
Front Panel Menu Reference SCPI Command
From the Calibration menu, select Date. Enter the date, then select Insert Date.
Then select Close.
From the Calibration menu, select Save. Select Save Calibration for all Outputs.
Then select Close.
From the Administrative Tools menu, select Administrator Login/Logout to log
out of the Admin menu.
To enter a calibration date:
CAL:DATE "<date>"
To savethe calibration data:
CAL:SAVE
To exitcalibration mode:
CAL:STAT OFF
8 Verification and Calibration
398 Keysight N6705C Operating and Service Guide