Peak Current Limit Calibration
For models N6753A through N6756A and N6763A through N6766A, the peak current
limit calibration is valid for five years.
Step 1. Connect only the electronic load or equivalent resistor (up to 500 W rating) across the output
terminals for the first calibration point.
Model 1st calibration point 2nd calibration point
N6751A/N6752A
N6761A/N6762A
25 Ω resistor (do notuse electronic load) 4 Ω resistor (do notuse electronic load)
N6753A/N6763A Electronic load set to 26.25A, CC mode (0.6 Ω equi-
valent, including lead resistance)
Electronic load set to 13.12A, CC mode (1.2 Ω equi-
valent, including lead resistance)
N6754A/N6764A Electronic load set to 10.5A, CC mode (3.8 Ω equi-
valent, including lead resistance)
Electronic load set to 5.25A, CC mode (7.6 Ω equi-
valent, including lead resistance)
N6755A/N6765A Electronic load set to 43.75A, CC mode (0.37 Ω equi-
valent, including lead resistance)
Electronic load set to 21.88A, CC mode (0.73Ω equi-
valent, including lead resistance)
N6756A/N6766A Electronic load set to 17.5A, CC mode (2.3 Ω equi-
valent, including lead resistance)
Electronic load set to 8.75A, CC mode (8.75Ω equi-
valent, including lead resistance)
Step 2. Select peak current limit calibration.
Front Panel Menu Reference SCPI Command
From the Calibration menu, select Miscellaneous, Peak Current Limit. Then
select Start.
CAL:CURR:PEAK (@1)
Step 3. Select the first current calibration point.
Front Panel Menu Reference SCPI Command
Select Next. The calibration is automatic. CAL:LEV P1
*OPC?
Step 4. Set the electronic load or connect the appropriate resistor for the second calibration point.
Step 5. Select the second current calibration point.
Front Panel Menu Reference SCPI Command
Select Next. The calibration is automatic. CAL:LEV P2
*OPC?
Downprogrammer Calibration
Step 1. Remove all loads from the output. This procedure is automatic and takes a few seconds.
Keysight N6705C Operating and Service Guide 397