CONFigure:CONTinuity
Sets all measurement parameters and trigger parameters to their default values for continuity meas-
urements.
Parameter Typical Return
(none) +1.32130000E-02
Configure the instrument for continuity measurements.Then make a measurement using an external trig-
ger with positive slope (rising edge) and read the measurement:
CONF:CONT
TRIG:SOUR EXT;SLOP POS
READ?
l The range and resolution are fixed at 1 kΩ for continuity tests (a 2-wire resistance measurement).
l The instrument beeps (if the beeper is enabled) for each measurement less than or equal to the con-
tinuity threshold (≤10 Ω), and the actual resistance measurement appears on the display.
l From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above
1.2 kΩ, the instrument displays "OPEN" with no beep.
l The FETCh?, READ?, and MEASure:CONTinuity? queries return the measured resistance, regardless of
its value.
l Use READ? or INITiate to start the measurement.
SCPI Programming Reference
246
Keysight Truevolt Series Operating and Service Guide