Finishing Calibration
To finish calibrating your instrument:
1.
Remove all shorting blocks and connections from the instrument.
2.
Reset the calibration message.
3.
Secure the instrument with the calibration password.
4.
Record the new calibration count.
5.
Perform the performance verification tests to verify the calibration.
Calibration Adjustment Procedures
554
Keysight Truevolt Series Operating and Service Guide