EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #550 background imageLoading...
Page #550 background image
AC Current 10 A Gain Calibration
This calibration procedure applies to the 34461A, 34465A, and 34470A only (not applicable to the
34460A).
Review Test Considerations and Flatness Calibration Considerations before performing any calibration.
Configuration: AC Current
All AC calibration procedures use the 3 Hz bandwidth measurement filter.
1.
Select the 10 A range.
2.
Apply the 5 A input signal at 1000 Hz.
3.
Enter the actual applied input current as the Cal Value (see Entering Calibration Values).A Calibration
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4.
Store the new calibration constants (see Storing the Calibration Constants).
This calibration takes approximately 2.5 seconds to complete.
Next Procedure
Keysight Truevolt Series Operating and Service Guide
551
Calibration Adjustment Procedures

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals