DATA Subsystem
This subsystem allows you to configure and remove data from reading memory. The instrument clears all
measurements from reading memory when the measurement configuration changes, or when any of these
commands are executed: INITiate, MEASure:<function>?, READ?, *RST, SYSTem:PRESet.
Command Summary
DATA:LAST?
DATA:POINts?
DATA:POINts:EVENt:THReshold
DATA:REMove?
DATA:LAST?
Returns the last measurement taken. You can execute this query at any time, even during a series of meas-
urements.
Parameter Typical Return
(none) (none)
Return the last measurement:
DATA:LAST?
Typical Response: One measurement with units. If no data is available, 9.91E37 (Not a Number) is returned
with units.
Example: +1.73730000E+00 VDC
DATA:POINts?
Returns the total number of measurements currently in reading memory. You can execute this query at
any time, even during a series of measurements.
Parameter Typical Return
(none) (none)
Return the number of measurements in reading memory:
DATA:POIN?
Typical Response: +215
l You can store up to 1,000 measurements in the reading memory of the 34460A, 10,000 meas-
urements on the 34461A, 50,000 measurements on the 34465A/70A (without the MEM option), or
2,000,000 measurements on the 34465A/70A (with the MEM option).
SCPI Programming Reference
256
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals