SYSTem:PRESet
This command is nearly identical to *RST. The difference is that *RST resets the instrument for SCPI oper-
ation, and SYSTem:PRESet resets the instrument for front panel operation. As a result, *RST turns the his-
togram and statistics off, and SYSTem:PRESet turns them on.
SYSTem:SECurity:COUNt?
Returns the instrument's security count.
Parameter Typical Return
(none) +22
Return the instrument's security count:
SYSTem:SECurity:COUNt?
l The security count is incremented by 1 every time you perform an action, other than calibration, that
requires the instrument to be unsecured. This includes enabling or disabling interfaces, sanitizing
memory per the NISPOM standard (SYSTem:SECurity:IMMediate), updating firmware, changing the
security code, or overriding the calibration security.
l The act of securing or unsecuring the instrument does not increment the counter.
l Be sure to read the security count upon receiving your instrument.
See Also
CALibration:COUNt?
Keysight Truevolt Series Operating and Service Guide
401
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals