4.
Reset the security code to the factory default:
a. Press [Shift] > [Utility] > Test / Admin > Security > New Code.
b.
Enter AT3446XA as the new passcode.
c. Press Done.
d. Verify the new code and press Yes to confirm the change.
5.
Remove power, disassemble the instrument, and remove the short. Then reassemble the instrument
to complete the security code (calibration password) reset procedure.
Performance Verification
Use the performance verification tests in this section to verify the measurement performance of the instru-
ment using the instrument’s specifications listed in the product data sheet.
You can perform three different levels of performance verification tests:
l
Self–Test (two types): Internal verification tests that give a high confidence that the instrument is
operational, as described here.
l
Quick Verification: A combination of the internal self–tests and selected verification tests.
l
Performance Verification Tests: An extensive set of tests that are recommended as an acceptance
test when you first receive the instrument or after performing calibration:
l Zero Offset Verification
l DC Volts and DC Current Gain Verification
l Frequency Accuracy Verification
l AC Voltage and AC Current Verification
l High Current Verification
l Capacitance Verification (Optional Verification Test)
Keysight Truevolt Series Operating and Service Guide
503
Performance Verification

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals