Resistance (2-wire) Default Delays
PLC
0.001
1
0.002
1
0.006
1
0.02
0.06
2
0.2 1 10 100
Aperture (60 Hz Power)
20 µs
1
40 µs
1
100 µs
1
0.3 ms
1 ms
2
3 ms
16.7 ms 0.167 s 1.67 s
Aperture (50 Hz Power) 20 ms 0.2 s 2 s
Range Delay
100 Ω 80 µs 100 µs 130 µs
1 kΩ 110 µs 130 µs 160 µs
10 kΩ 130 µs 160 µs 190 µs
100 kΩ 540 µs 670 µs 800 µs
1 MΩ 5 ms 6 ms 7.5 ms
10 MΩ
60 ms 70 ms 84 ms
100 MΩ
1 GΩ
2
1
34465A/70A with DIG option only.
2
34465A/70A only.
Resistance (4-wire) Default Delays
PLC
0.001
1
0.002
1
0.006
1
0.02
0.06
2
0.2 1 10 100
Aperture (60 Hz Power)
20 µs
1
40 µs
1
100 µs
1
0.3 ms
1 ms
2
3 ms
16.7 ms 0.167 s 1.67 s
Aperture (50 Hz Power) 20 ms 0.2 s 2 s
Range Delay
100 Ω
1 ms
1.5 ms
1 kΩ
10 kΩ
100 kΩ
1 MΩ 10 ms 15 ms
10 MΩ
100 ms
100 MΩ
1 GΩ
2
1
34465A/70A with DIG option only.
2
34465A/70A only.
AC Voltage Default Delays
Filter (Hz)
Range 3 20 200
All 2.5 s 0.625 s 0.025 s
Keysight Truevolt Series Operating and Service Guide
457
SCPI Programming Reference