EasyManua.ls Logo

Keysight Truevolt Series - Resistance (2-Wire) Default Delays

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Resistance (2-wire) Default Delays
PLC
0.001
1
0.002
1
0.006
1
0.02
0.06
2
0.2 1 10 100
Aperture (60 Hz Power)
20 µs
1
40 µs
1
100 µs
1
0.3 ms
1 ms
2
3 ms
16.7 ms 0.167 s 1.67 s
Aperture (50 Hz Power) 20 ms 0.2 s 2 s
Range Delay
100 80 µs 100 µs 130 µs
1 kΩ 110 µs 130 µs 160 µs
10 kΩ 130 µs 160 µs 190 µs
100 kΩ 540 µs 670 µs 800 µs
1 M 5 ms 6 ms 7.5 ms
10 M
60 ms 70 ms 84 ms
100 M
1 GΩ
2
1
34465A/70A with DIG option only.
2
34465A/70A only.
Resistance (4-wire) Default Delays
PLC
0.001
1
0.002
1
0.006
1
0.02
0.06
2
0.2 1 10 100
Aperture (60 Hz Power)
20 µs
1
40 µs
1
100 µs
1
0.3 ms
1 ms
2
3 ms
16.7 ms 0.167 s 1.67 s
Aperture (50 Hz Power) 20 ms 0.2 s 2 s
Range Delay
100
1 ms
1.5 ms
1 kΩ
10 kΩ
100 kΩ
1 M 10 ms 15 ms
10 M
100 ms
100 M
1 GΩ
2
1
34465A/70A with DIG option only.
2
34465A/70A only.
AC Voltage Default Delays
Filter (Hz)
Range 3 20 200
All 2.5 s 0.625 s 0.025 s
Keysight Truevolt Series Operating and Service Guide
457
SCPI Programming Reference

Table of Contents

Related product manuals