SCPI Error Messages
Command Errors (-100…)
Execution Errors (-200…)
Device-Specific Errors (-300…)
Query Errors (-400…)
Network Errors (+100...)
Instrument Errors (+200…)
Miscellaneous Errors (+300... and +500..)
Licensing and Self-test Errors (+600...)
Calibration Errors (+700…)
Miscellaneous Errors (+800...)
The instrument returns error messages in accordance with the SCPI standard.
l Up to 20 errors can be stored in the instrument's error queue. Each remote interface I/O session
(GPIB, USB, VXI-11, Telnet/Sockets) has its own interface-specific error queue. Errors appear in the
error queue of the I/O session that caused the error. For example, if an error was generated by a com-
mand sent over GPIB, send SYSTem:ERRor? from GPIB to read the error queue. Errors generated by
instrument hardware are broadcast to all I/O sessions.
l The instrument beeps each time an error is generated (unless disabled by SYSTem:BEEPer:STATe OFF).
The front panel Error annunciator turns on when one or more errors are in the error queue.
l A special global error queue holds all power-on and hardware-related errors (for example, over-
temperature).
l Error retrieval is first-in-first-out (FIFO), and errors are cleared as you read them. When you have read
all of the interface-specific errors, the errors in the global error queue are retrieved. When you have
read all errors from the interface-specific and global error queues, the Error annunciator turns off.
l If more than 20 errors have occurred, the most recent error stored in the queue is replaced with -
350,"Queue overflow". No additional errors are stored until you remove errors from the queue. If no
errors have occurred when you read the error queue, the instrument responds with +0,"No error".
l The front panel reports errors from all I/O sessions and the global error queue. To read the error queue
from the front panel, press Help > View remote command error queue.
l Error conditions are also summarized in the Status Byte Register. For more information on the SCPI
Status System, see Status Subsystem Introduction.
l The interface-specific error queues are cleared by the *CLS (Clear Status) command and when power is
cycled. All errors are cleared when you read the error queue. The error queue is not cleared by a Fact-
ory Reset (*RST) or an Instrument Preset (SYSTem:PRESet).
SCPI Programming Reference
460
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals