CONFigure Subsystem
The CONFigure commands are the most concise way to configure measurements. Like the MEASure? quer-
ies, these commands use default measurement configuration values. However, these commands do not
automatically start measurements, so you can modify measurement attributes before initiating the meas-
urement.
Use INITiate or READ? to initiate measurements.
Command Summary
CONFigure?
CONFigure:CAPacitance
CONFigure:CONTinuity
CONFigure:CURRent:{AC|DC}
CONFigure:DIODe
CONFigure:{FREQuency|PERiod}
CONFigure:{RESistance|FRESistance}
CONFigure:TEMPerature
CONFigure[:VOLTage]:{AC|DC}
CONFigure[:VOLTage][:DC]:RATio
Default Settings for the CONFigure Command
The CONFigure commands select the function, range and resolution in one command. Specify
<resolution> in the measurement's units (V, A, Hz, Ω, and so on). All other parameters are set to their
default values (below).
Measurement Parameter Default Setting
AC Input Filter (bandwidth) 20 Hz (medium filter)
Autozero OFF if resolution setting results in NPLC < 1
ON if resolution setting results in NPLC ≥ 1
Range AUTO (including voltage range for frequency and period measurements)
Samples per Trigger 1 sample
Trigger Count 1 trigger
Trigger Delay Automatic delay
Trigger Source Immediate
Trigger Slope NEGative
Math Functions Disabled. Other parameters are unchanged.
Per-function Null State Disabled
Keysight Truevolt Series Operating and Service Guide
243
SCPI Programming Reference