EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #358 background imageLoading...
Page #358 background image
[SENSe:]TEMPerature:APERture {<seconds>|MIN|MAX|DEF}
[SENSe:]TEMPerature:APERture? [{MIN|MAX|DEF}]
Specifies the integration time in seconds (called aperture time) for temperature measurements.
Use this command for precise control of the DMM's integration time. Use NPLC (see
[SENSe:]TEMPerature:NPLC) for better power-line noise rejection characteristics
(NPLC > 1).
Parameter Typical Return
(Without the DIGoption) 200µs to 1 s (2 µs resolution), default: 100 ms. (With the DIG
option) 20 µs to 1 s (2 µs resolution), default: 100 ms.
+1.00000000E-
01
Enable aperture, set the aperture time to 300ms:
TEMP:APER:ENAB ON
TEMP:APER 300E-03
l Applies only to the 34465A and 34470A.
l
Both the [SENSe:]<function>:NPLC and [SENSe:]<function>:APERture commands control the instru-
ment's integration time. When using both commands (not recommended), whichever command is
sent last will override the other command and set the integration time. Sending the
[SENSe:]<function>:APERture command sets the [SENSe:]<function>:APERture:ENABled command
to 1 (ON); sending [SENSe:]<function>:NPLC sets it to 0 (OFF).
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
[SENSe:]TEMPerature:APERture:ENABled {ON|1|OFF|0}
[SENSe:]TEMPerature:APERture:ENABled?
Enables the setting of integration time in seconds (called aperture time) for temperature measurements.
If aperture time mode is disabled (default), the integration time is set in PLC (power-line cycles).
Parameter Typical Return
{ON|1|OFF|0}. Default: OFF. 0(OFF)or1(ON)
Enable aperture time mode, set the aperture time to 300ms:
TEMP:APER:ENAB ON
TEMP:APER 300E-03
l Applies only to the 34465A and 34470A.
l The CONFigure:TEMPerature, MEASure:TEMPerature?,and [SENSe:]TEMPerature:NPLC commands
disable the aperture time mode, and select an integration time in number of power line cycles.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
359
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals