Calibration Message
You can store a message of up to 40 characters in calibration memory. For example, you could store the
date of the last calibration, the date the next calibration is due, or the name and phone number of the per-
son responsible for calibration.
To read the calibration message, press Utility > Test/Admin > Calibrate or send CALibration:STRing?
from the remote interface.
To change the calibration message, unsecure the instrument and press Utility > Test/Admin > Calibrate
> Edit Cal String or send CALibration:STRing "<string>" from the remote interface.
Calibration Adjustment Procedures
532
Keysight Truevolt Series Operating and Service Guide