LXI Subsystem
This subsystem supports LAN eXtensions for Instrumentation (LXI) functionality.
Command Summary
LXI:IDENtify[:STATe]
LXI:MDNS:ENABle
LXI:MDNS:HNAMe[:RESolved]?
LXI:MDNS:SNAMe:DESired
LXI:MDNS:SNAMe[:RESolved]?
LXI:RESet
LXI:RESTart
LXI:IDENtify[:STATe] {ON|1|OFF|0}
LXI:IDENtify[:STATe]?
Shows or removes the LXI Web Identify indicator on the display.
Parameter Typical Return
{ON|1|OFF|0}. Default: OFF. 0(OFF)or1(ON)
Turn on the LXI Web Identify indicator:
LXI:IDEN ON
l The indicator helps you identify the device associated with the LAN address.
l
Press the [Local] key or send *RST to turn off the indicator.
LXI:MDNS:ENABle {ON|1|OFF|0}
LXI:MDNS:ENABle?
Disables or enables the multicast Domain Name System (mDNS), which provides the capabilities of a DNS
server for service discovery in a small network without a DNS server.
Parameter Typical Return
{ON|1|OFF|0}. Default: ON. 0(OFF)or1(ON)
Turn mDNS ON:
LXI:MDSN:ENAB ON
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru-
ment Preset (SYSTem:PRESet).
l This parameter is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.
Keysight Truevolt Series Operating and Service Guide
277
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals