[SENSe:]{RESistance|FRESistance}:NULL:VALue {<value>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for all resistance measurements.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter Typical Return
-120 MΩ to +120 MΩ. Default: 0. (34460A/61A)
-1.2 GΩ to +1.2 GΩ. Default: 0. (34465A/70A)
+1.04530000E+02
Configure 2-wire resistance measurements, using the null function to remove 100 mΩ of wiring resistance.
Make and read two measurements:
CONF:RES
RES:NULL:STAT ON;VAL .1
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
l Specifying a null value disables automatic null value selection ([SENSe:]
{RESistance|FRESistance}:NULL:VALue:AUTO OFF).
l To use the null value, the null state must be on ([SENSe:]{RESistance|FRESistance}:NULL:STATe ON).
l This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset
(SYSTem:PRESet), or CONFigure function.