Using Device Clear
Device Clear is an IEEE-488 low-level bus message that you can use to return the instrument to a respons-
ive state. Different programming languages and IEEE-488 interface cards provide access to this capability
through their own unique commands. The status registers, the error queue, and all configuration states
are left unchanged when a Device Clear message is received.
Device Clear performs the following actions:
l If a measurement is in progress, it is aborted.
l The instrument returns to the trigger "idle" state.
l The instrument's input and output buffers are cleared.
l The instrument is prepared to accept a new command string.
l An overlapped command, if any, is terminated with no "Operation Complete" indication (applies to
INIT).
The ABORt command is the recommended method to terminate a measurement.
Commands by Subsystem
ABORt
CALCulate:CLEar[:IMMediate]
CALCulate:AVERage Subsystem
CALCulate:LIMit Subsystem
CALCulate:SCALe Subsystem
CALCulate:SMOothing Subsystem
CALCulate:TCHart[:STATe]
CALCulate:TRANsform:HISTogram Subsystem
CALibration Subsystem
CONFigure Subsystem
DATA Subsystem
DISPlay Subsystem
FETCh?
FORMat Subsystem
HCOPy Subsystem
IEEE 488.2 Common Commands
INITiate[:IMMediate]
LXI Subsystem
Keysight Truevolt Series Operating and Service Guide
199
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals