Recommended Test Equipment
The recommended test equipment for the performance verification and calibration is listed below. If the
exact instrument is not available, substitute calibration standards of equivalent accuracy.
An alternate method is to use the Keysight 3458A 8½-digit Digital Multimeter to measure less accurate,
yet stable, sources. The output value measured from the source can be entered into the instrument as the
target calibration value.
Application Recommended Equipment
Zero Calibration Keysight 34172B
DC Voltage Fluke 5720A
DC Current Fluke 5720A
Resistance Fluke 5720A
AC Voltage Fluke 5720A
AC Current Fluke 5720A
Frequency Keysight 33500B Series Function/Arbitrary Waveform Generator
High Current Fluke 5725A
Capacitance (Optional) SCA-1uF Capacitance Standard
Calibration Adjustment Procedures
526
Keysight Truevolt Series Operating and Service Guide