EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
Additional Digitize Default Settings
On entering digitize mode, the DMM configures these settings:
l If trigger source is set to manual it is changed to auto. (external and level remain as is.)
l Limits mode is turned off.
l Scaling is turned off.
l Statistic and histogram are put in post-processing mode (computed after digitize is complete).
l On the selected function (DCV or DCI) and for the new function if changed:
l Autorange is turned off.
l Autozero is turned off.
l NPLC and aperture are set to their minimum values.
l If trigger source is external or level, the pretrigger count is set to the digitize pretrigger count setting
(default of 0).
l Samples per trigger is set according to the digitize duration (time or samples).
l Sample timer is put in timer (not immediate) mode and Sample timer is set according to the digitize
sample rate or sample interval.
l Trend Chart mode is changed to bucketized when data logging to a file.
l On return to Continuous mode, settings are left as done in data log or digitize mode except:
l Sample source is set to immediate.
l Pretrigger count is set to 0.
l Samples per trigger is set to 1.
l Trigger count is set to infinite.
Keysight Truevolt Series Operating and Service Guide
49
Features and Functions

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals