EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #279 background imageLoading...
Page #279 background image
MEASure Subsystem
The MEASure queries are the easiest way to program measurements because they always use default meas-
urement parameters. You set the function, range, and resolution in one command, but you cannot change
other parameters from their default values. The results are sent directly to the instrument's output buffer.
A MEASure query is functionally equivalent to sending CONFigure followed immediately by READ?. The
difference is that CONFigure commands allow you to change parameters between the CONFigure and the
READ?
Command Summary
MEASure:CAPacitance?
MEASure:CONTinuity?
MEASure:CURRent:{AC|DC}?
MEASure:DIODe?
MEASure:{FREQuency|PERiod}?
MEASure:{RESistance|FRESistance}?
MEASure:TEMPerature?
MEASure[:VOLTage]:{AC|DC}?
MEASure[:VOLTage][:DC]:RATio?
Default Settings for MEASure?
With the MEASure? queries, you can select the function, range and resolution in one command. All other
parameters are set to their default values (below).
Measurement Parameter Default Setting
AC Input Filter (bandwidth) 20 Hz (medium filter)
Autozero OFF if resolution setting results in NPLC < 1
ON if resolution setting results in NPLC ≥ 1
Range AUTO (including voltage range for frequency and period measurements)
Samples per Trigger 1 sample
Trigger Count 1 trigger
Trigger Delay Automatic delay
Trigger Source Immediate
Trigger Slope NEGative
Math Functions Disabled. Other parameters are unchanged.
Per-function Null State Disabled
SCPI Programming Reference
280
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals