EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #287 background imageLoading...
Page #287 background image
MEASure:TEMPerature? [{FRTD|RTD|FTHermistor|THERmistor|TCouple|DEFault} [, {<type>|DEFault} [,1
[, {<resolution>|MIN|MAX|DEF}]]]]
Sets all measurement parameters and trigger parameters to their default values for temperature meas-
urements and immediately triggers a measurement. The results are sent directly to the instrument's out-
put buffer.
Parameter Typical Return
<probe_type>: {FRTD|RTD|FTHermistor|THERmistor|TCouple}. Default: FRTD.
(none)
<type>: 85 (only possible value for RTD/FRTD), 5000 (only possible value for THERmis-
tor/FTHermistor), or E, J, K, N, R, T (TCouple).
<resolution>: See Resolution Table or Range, Resolution and NPLC. The
default is equivalent to 10 PLC.
Configure a 4-wire RTD measurement with default resolution. Then make and read one measurement:
MEAS:TEMP? FRTD,85
Typical Response:+2.12320000E+01
l Thermocouple measurements (TC) apply only to the 34465A and 34470A.
l For temperature measurements, the instrument internally selects the range; you cannot select the
range to be used. The range ("1" is the only allowed value) is used only in conjunction with resolution
to determine NPLC.
l
The <resolution>parameter only determines the integration time; it does not specify the resolution of
the temperature measurement. This parameter is optional; however, if you specify <resolution>, you
must also specify "1" as the implied range parameter. For example: CONF:TEMP
RTD,85,1,0.000001 selects the 10 PLC integration time on the 34461A.
l To change temperature units, use UNIT:TEMPerature.
l For RTD and thermistor measurements, the instrument autoranges to the correct range for the trans-
ducer resistance measurement. For thermocouple measurements, the 100 mV range is selected.
l For thermocouple measurements, the internal reference is selected. See SENSe:TEMPer-
ature:TCouple:RJUNction:TYPE.
l If the input signal is greater than can be measured on the specified manual range, the instrument dis-
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
SCPI Programming Reference
288
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals