EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #208 background imageLoading...
Page #208 background image
UNIT:TEMPerature {C|F|K}
UNIT:TEMPerature?
Selects the units (°C, °F or Kelvin) to be used for all temperature measurements except the
SYSTem:TEMPerature? query.
Parameter Typical Return
{C|F|K}. Default: C. C, F, or K
Make a four-wire RTD measurement and return the result in °F:
UNIT:TEMP F
MEAS:TEMP? FRTD
Typical Response: +6.82320000E+01
l The command also accepts CEL or FAR, but the query returns C or F.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
See Also
CONFigure:TEMPerature
MEASure:TEMPerature?
[SENSe:]FUNCtion[:ON]
Keysight Truevolt Series Operating and Service Guide
209
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals