SYSTem:TIME
SYSTem:DATE
CALibration:VALue <value>
CALibration:VALue?
Specifies the value of the applied calibration signal.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Parameter Typical Return
Numeric,default0.0 +2.37000000E-02
Specify calibration value 0.0237:
CAL:VAL 2.37E-2
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
SCPI Programming Reference