width.
l Use READ? or INITiate to start the measurement.
See Also
CONFigure?
MEASure:CURRent:{AC|DC}?
[SENSe:]CURRent:AC:BANDwidth
CONFigure:DIODe
Sets all measurement parameters and trigger parameters to their default values for diode tests.
Parameter Typical Return
(none) (none)
Configure, make, and read a default diode measurement:
CONF:DIOD
READ?
Typical Response: +1.32130000E-01
l
The range and resolution are fixed for diode tests: the range is 1 VDC (with a 1 mA current source out-
put).
l The voltage is displayed on the front panel if it is between 0 and 5.05 V. The instrument beeps when
the signal transitions into the range between 0.3 and 0.8 V (unless beeper is disabled). If the signal
exceeds 5.05 V, the front panel shows "OPEN", and the value returned from SCPI is 9.9E37.
l The FETCh?, READ?, and MEASure:DIODe? queries return the measured voltage, regardless of its
value.
l Use READ? or INITiate to start the measurement.
SCPI Programming Reference
248
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals