Component side of board
For the following measurements, the instrument should be turned on, and the light below the power
switch should be green.
Refer to the detailed images in the table below for exact location information.
Pin Location Detail
Low reference: chassis
J1102 pin 21:
3.3 V ± 5%
Low reference: large metal shield
R1020 (either side of component):
10 V ± 20%
Low reference: large metal shield
E1006 (either side of component):
3.3 V ± 5%