EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #506 background imageLoading...
Page #506 background image
Values for 34461A DMM
Error from Nominal
Input Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
Open DC Current
100 µA Q ± 0.02 µA ± 0.025 µA
1 mA ± 0.060 µA
10 mA ± 2 µA
100 mA ± 4 µA ± 5 µA
1 A ± 60 µA ± 100 µA
3 A ± 600 µA
10 A ± 1 mA
Short DC Volts
100 mV Q ± 3 µV ± 3.5 µV
1 V ± 6 µV ± 7 µV
10 V ± 40 µV ± 50 µV
100 V ± 600 µV
1000 V ± 6 mV ± 10 mV
Short 4-wire Ohms
100 Ω ± 3 mΩ ± 4 mΩ
1 kΩ ± 5 mΩ ± 10 mΩ
10 kΩ Q ± 50 mΩ ± 100 mΩ
100 kΩ ± 500 mΩ ± 1 Ω
1 MΩ ± 10 Ω
10 MΩ ± 100 Ω
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Keysight Truevolt Series Operating and Service Guide
507
Performance Verification

Table of Contents