Values for 34470A DMM
Error from Nominal
Input Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
Open DC Current
1 µA ± 50 pA (50E-12 A)
[2]
10 µA ± 200 pA 200.0E-12 A)
[2]
100 µA Q ± 0.001 µA
[2]
1 mA ± 0.030 µA ± 0.050 µA
10 mA ± 2 µA
100 mA ± 4 µA ± 5 µA
1 A ± 60 µA ± 100 µA
3 A ± 600 µA
10 A ± 1 ma
Short DC Volts
100 mV Q ± 3 µV ± 3.5 µV
1 V ± 4 µV
10 V ± 20 µV
100 V ± 600 µV
1000 V ± 6 mV
Short 4-wire Ohms
100 ± 3 mΩ ± 4 m
1 kΩ ± 5 mΩ
10 kΩ Q ± 50 mΩ
100 kΩ ± 500
1 MΩ ± 5 Ω
10 MΩ ± 100 Ω
Short 2-wire Ohms 100 1
[1]
Select 10 PLC.
[2]
These values are typical and these ranges are calibrated internally.
Q: Quick performance verification test point.
Keysight Truevolt Series Operating and Service Guide