are recorded as usual. If the trigger event occurs before Pretrigger Count readings have been taken, the
trigger event will still take effect and digitizing will complete without having taken all of the pretrigger read-
ings. The Pretrigger Count is limited to one less than the total number of readings specified to be taken by
the Duration setting.
Step 8: Press [Run/Stop] to start the digitizing process. Digitizing will begin when the specified trigger
event occurs and after the specified delay has elapsed. • Digitizing is displayed at the top of the display
while digitizing; when finished, Digitize Stopped is displayed.
Step 9: All readings taken during digitizing are saved in volatile memory, when digitizing is finished, press
the Save Readings softkey to specify a file location and save the readings.
Step 10: (Optional) Press Display > Display Trend to view the trend chart. The Trend Chart is particularly
useful for viewing and analyzing digitized measurements. X and Y cursors (shown below), tracking cursors,
pan control, and zoom allow you to analyze digitized data. See Trend Chart (Digitize and Data Log Modes)
for details.
Features and Functions
90
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals