Clearing Reading Memory
The following actions clear reading memory:
l Changing the measurement function
l
Pressing any Clear Readings softkey
l Transitioning in or out of Probe Hold
l Changing temperature units
l Changing any dB/dBm parameters
l Changing any histogram binning parameter
l Changing the temperature probe or R
0
l Recalling a stored state
l Calibrating the instrument
l Switching between 3 A and 10 A inputs
l Changing the position of the Front/Rear switch
l Transferring from Remote to Local mode
l Turning Null on or off or changing the Null value
These actions do not clear reading memory:
l Changing measurement parameters, such as range and aperture.
l Turning limits on or off, or adjusting limit values
l
Pressing [Run/Stop] in front panel auto trigger mode
l
Changing the trend chart Recent/All softkey (34461A/65A/70A only)
l Changing samples per trigger or trigger delay
l Changing display modes
l
Changing the VM Comp output polarity
l Changing the digit mask
l Changing the histogram, bar meter, or trend chart scale
l Changing user preferences
l Executing self-test
Keysight Truevolt Series Operating and Service Guide
99
Features and Functions