TOS5300 65
SIGNAL I/O Connector
6
External Control
■ SIGNAL I/O connector pinout
Pin no. Signal name I/O TOS5300 TOS5301 TOS5302
1
INTERLOCK+
I
If you open the positive and negative terminals, the output is turned off, and the TOS5300 Series is
switched to Protection mode.
Open: The resistance between the two terminals is 1.2 kΩ or greater.
Short: The resistance between the two terminals is 1 kΩ or less.
2
PM0
I
Panel memory selection signal.
The selection signal is latched on the
rising edge of the input strobe signal to
recall panel memory.
* The selection of memory is prioritized
over TEST SEL and AUTO SEL.
3
PM1
I
4NC ——
5NC ——
6NC ——
7NC ——
8NC ——
9 STB I Panel memory strobe signal input terminal.
10
TEST SEL
I
NA ACW/DCW selection signal
input.
L: ACW. H: DCW.
Single test selection signal and auto test
order selection signal.
If AUTO SEL specifies single test, L: ACW
and H: IR.
If AUTO SEL specifies auto operation, L:
ACW→
IR and H: IR→ACW.
11
AUTO SEL
I
NA NA Auto test/single test selection.
L: Single test. H: Auto test.
12 COM — Circuit common terminal.
13
INTERLOCK-
I
If you open the positive and negative terminals, the output is turned off, and the TOS5300 Series is
switched to Protection mode.
Open: The resistance between the two terminals is 1.2 kΩ or greater. Short: The resistance between the
two terminals is 1 kΩ or less.
14
HV.ON
O
On during testing and when a voltage remains across the
output terminals.
On during testing, when a voltage
remains across the output terminals,
and during auto tests (AUTO TEST).
15 TEST O On during testing (excluding when voltage is rising or falling).
16 PASS O
On for at least 0.2 seconds (the PASS HOLD time) when a PASS judgment occurs.
On continuously when the PASS HOLD time is set to HOLD.
17 U-FAIL O
On continuously when an UPPER FAIL judgment occurs because a value greater than or equal to the
upper limit is detected.
18 L-FAIL O
On continuously when a LOWER FAIL judgment occurs because a value less than or equal to the lower
limit is detected.
19 READY O On when the TOS5300 Series is waiting (when it is in the READY state).
20 PROTECTION O On when protective features have been activated.
21 START I Start signal input terminal.
22 STOP I Stop signal input terminal.
23 ENABLE I
Input terminal for the start signal’s ENABLE signal. If the ENABLE signal changes, the TOS5300 Series is
switched to Protection mode.
24 +24 V — +24 V internal power supply output terminal; maximum output current 100 mA.
25 COM — Circuit common terminal.
PM0 PM1 Recalled panel memory number
HHMemory 1
L H Memory 2
H L Memory 3
L L Enables the selection of TEST SEL and AUTO SEL
9
8
7
6
5
4
3
2
1
13
12
11
10
19
18
17
16
15
14
24
23
22
21
20
25
NC
NC
NC
NC
NC
PM1
PM0
INTERLOCK+
INTERLOCK-
COM
AUTO SEL
TEST SEL
STB
COM
+24V
ENABLE
STOP
START
PROTECTION
READY
L-FAIL
U-FAIL
PASS
TEST
HV.ON