TOS5300 95
Troubleshooting
App
Unable to perform measurements
Symptom Items to check and possible remedy
The measured values are abnormal. Are the test leads connected to the correct terminals?
Connect the test leads properly.
p.26
Judgments do not begin. Has the judgment wait time been set?
Judgments will not begin until the specified judgment wait time elapses. Set an
appropriate judgment wait time.
p.53
Testing does not finish. Is the timer turned off?
The specified test time has not been enabled. Turn the timer on.
p.46
p.47
p.48
p.49
The time when test data was saved is
incorrect.
Is the system clock correct?
Set the system clock to match the current time.
p.73
See