6.6.7 Calibration Mode: Temperature
Adjustment of Temperature Probe
This function is used to adjust the individual temperature probe tolerances or cable lengths for the
purpose of increasing the accuracy of the temperature measurement.
The adjustment requires an accurate measurement of the process temperature using a calibrated
reference thermometer. The measurement error of the reference thermometer should be less than 0.1
K. Adjustment without an accurate measurement may result in falsification of the displayed measured
value.
When using Memosens sensors, the adjustment value is saved in the sensor.
Calibration Procedure
Calibration
[I] [II] [Sensor]
Adjust Temperature Probe
01. Select Temperature calibration mode and press enter to confirm.
02. Enter the measured process temperature and press enter to confirm.
The temperature offset is displayed.
03. Adjust the temperature probe with the right softkey: Save.