EasyManua.ls Logo

Knick Stratos Multi E401N - pH

Knick Stratos Multi E401N
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Stratos Multi E401N
159
12.8 Measuring Functions
12.8.1 pH
Digital input
For Memosens sensors (pH, ORP, pH/ORP)
Terminals 1…5 or MK-MS095N module
Display ranges Temperature: -20.0…200.0°C / -4…392°F
pH value: -2.00…16.00
ORP: -1999 ... 1999mV
rH value (with pH/ORP sensor): 0…42.5
Measurement error Depending on sensor
Module input, analog
For analog pH and ORP sensors
Measuring ranges Temperature: -20.0…200.0°C / -4…392°F
pH value: -2.00…16.00
ORP: -1999 ... 1999mV
rH value (with pH/ORP sensor): 0…42.5
Glass electrode input
Reference temperature 25°C/77°F
Input resistance > 1x10
12
Ω
Input current < 1 x 10
-12
A
Impedance range: 0.5...1000MΩ (±20%)
Reference electrode input
Reference temperature 25°C/77°F
Input resistance > 1x10
10
Ω
Input current < 1 x 10
-10
A
Impedance range: 0.5...200kΩ (±20%)
Measurement error
1)
2)
pH value < 0.02, TC: 0.002pH/K
mV value < 1mV, TC: 0.1mV/K
Temperature input via module
Pt100/Pt1000/NTC30kΩ/NTC 8.55kΩ/Balco 3kΩ
2-wire connection, adjustable
Measuring ranges Pt100/Pt1000: -20.0 ... 200.0°C/-4 ... 392°F
NTC30kΩ: -20.0 ... 150.0°C/-4 ... 302°F
NTC8.55kΩ (Mitsubishi): -10.0...130.0°C/14...266°F
Balco 3kΩ: -20.0...130.0°C/-4...266°F
Adjustment range 10K
Resolution 0.1°C/0.1°F
Measurement error
1)
2)
< 0.5K (< 1K for Pt100; <1K for NTC >100°C/212°F)
Temperature compensation
Off,
Linear characteristic 00.00...19.99%/K,
Ultrapure water,
Table: 0 … 95°C, user-defined in 5-K steps
Reference temperature 25°C/77°F
1)
At rated operating conditions
2)
± 1 count, plus sensor error

Table of Contents