EasyManua.ls Logo

Knick Stratos Multi E401N - Calibration Presettings; Pressure Correction

Knick Stratos Multi E401N
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Stratos Multi E401N
95
5.12.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the
Calibration menu prior to the calibration.
Calibration mode: Presetting of calibration mode, e.g., in air, in water, data entry, product calibration,
zero calibration, temperature
If you select “Product Calibration mode, you must also select the measured value: Saturation %air,
concentration (gas), partial pressure
Calibration timer: When a preset calibration interval expires, the calibration timer generates a mes-
sage text to indicate the need for calibration. If Auto is selected, the interval is set to 720 h.
If “Individual” is selected, you can select a custom interval.
Cal Presettings (Admin.)
Product Calibration
Back
Calibration Mode
Calibration Timer
Product
Saturation
II
Sat. %Air
Conc. (Liquid)
Partial Pressure
Settings can be changed in the Cal Presettings submenu:
Parameter Setting
[I] [II] … Oxy
Cal Presettings
Note: If Sensoface is enabled, a neutral smiley is displayed once 80% of the interval has expired. Once
the entire interval has expired, a sad smiley is shown, a maintenance required message is generated,
and the corresponding NAMUR icon is displayed. If the current outputs are configured accordingly,
a 22-mA error signal is generated.
5.12.3 Pressure Correction
The pressure used for measurement or calibration can be manually specified (factory setting
1013mbar).
With TAN option FW-E051, “Current Input”, an external pressure transmitter can be connected to the
current input (terminals 8 and 9). This enables automatic pressure correction. The start and end of the
current input can be set in the ranges from 0/4 ... 20mA.
Settings can be changed in the Pressure Correction submenu:
Parameter Setting
[I] [II] ... Oxy
Pressure Correction
Setting Automatic Pressure Correction
(TAN Option FW-E051)
01. Open the Ext. Pressure Transmitter submenu.
02. Select Absolute or “Differential” pressure transmitter.
03. Select current input 0 … 20 mA or 4 … 20mA.
04. Enter pressure values for current start and current end
05. Use the left softkey: Back to return to the Pressure Correction submenu.
06. Select an external or manual pressure correction in Pressure during Meas. and Pressure during Cal .

Table of Contents