Stratos Multi E401N
43
5.3 System Control
Submenus Description
Memory Card This option is shown if a Data Card is inserted:
Settings for logbook and measurement recorder data recording. Decimal separation
can be set. The memory card can be formatted.
Transfer Configuration If a Data Card is inserted, the measuring devices configuration can be saved and
transferred to another measuring device.
Transfer Configuration, p. 44
Parameter Sets Two parameter sets (A, B) are available in the device. If a Data Card is inserted, up to
five parameter sets can be saved on or loaded from the Data Card.
Parameter Sets, p. 45
Function Control Allocation of functions for activation by softkey or optocoupler input OK1.
Function Control, p. 46
Calculation Blocks TAN option FW-E020: Convert available process variables to new
variables.
Calculation Blocks (TAN Option FW-E020), p. 46.
Time/Date
Define date and time format; input of date, time, and weekday. Time/Date, p. 46
Meas. Point Description Free input of a tag number and notes; can be retrieved in the Diagnostics menu.
Measuring Point Description, p. 46
Firmware Update This option is shown if a FW Update Card is inserted. TAN option FW-E106: Firmware
update with FW Update Card.
Firmware Update (TAN Option FW-E106), p. 46
Enable Options Activation of add-on options via TAN. The TAN is only valid for the Stratos Multi with
the associated serial number.
Option Activation, p. 47
Logbook Select events to be logged (failure/maintenance required); can be retrieved in the Di-
agnostics menu.
Logbook, p. 47
Buffer Table TAN option FW-E002: Specification of a buffer set.
pH Buffer Table: Entry of Individual Buffer Set (FW-E002), p. 199
Concentration Table TAN option FW-E009: Specification of a concentration solution for conductivity mea-
surement.
Concentration Determination (FW-E009), p. 201.
Restore Factory Settings
Restore all parameters to factory settings. Restore Factory Settings, p. 48
Passcode Entry
Change passcodes. Passcode Entry, p. 48