EasyManua.ls Logo

Knick Stratos Multi E401N - Sensor Connection

Knick Stratos Multi E401N
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Stratos Multi E401N
31
3.3 Sensor Connection
3.3.1 Connecting a Memosens Sensor/Optical Oxygen Sensor (LDO)
Top view of terminals for Memosens/LDO sensor. The figure shows the opened device, rear side of the
front unit.
1
Standard sensor connection for digital sensors (Memosens sensor/LDO SE740 sensor)
2
Terminal plate with terminal assignments for digital sensor
3
Sensor connection for analog sensors or second Memosens sensor via measuring module
Note: The terminals are suitable for single or stranded wires up to 2.5mm
2
.
Memosens sensor SE 740 optical oxygen sensor (LDO)
Terminal Wire color Wiring
Memosens cable
Terminal Wire color Wiring
M12 cable
1 Brown (BN) +3V 1 -
2 Green (GN) RS-485 A 2 Gray RS-485 A
3 Yellow (YE) RS-485 B 3 Pink RS-485 B
4 White (WH) GND 4 Brown Shield
5 Transparent (CL) Shield 5 - -
6 6 White Power Out
01. Using an appropriate sensor cable, connect a Memosens sensor or the SE740 optical oxygen
sensor (LDO) to the RS-485 interface (2)
of the Stratos Multi.
02. Then select a measuring function and configure the sensor:
From within measuring mode, press the left softkey: Menu
.
The Menu selection opens.
03. Select Parameter Setting
Sensor Selection [I] [II] .
Note: Function check (HOLD) is active.
04. Press enter to open the Sensor Selection [I] .
05. Select process variable, mode, and functionality, and confirm with enter.
Set further parameters with the left softkey: Back
.
06. Return to measuring mode to exit parameter setting,
e.g., with the right softkey: Back to Meas
.

Table of Contents