Stratos Multi E401N
83
5.10 Process Variable Conductivity (Inductive)
Note: Function check (HOLD) is active.
Note: After changing the process variable or measuring mode, Stratos Multi retains its settings but
needs to be reconfigured.
Selecting a Digital Toroidal Conductivity Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [I]
Selection of toroidal Memosens conductivity sensor connected to the RS-485 interface
(terminals 1 … 5):
Parameter: Auto or Conductivity (Induct.)
Mode: Memosens
Functionality: CondI
Selection of an SE680 toroidal digital conductivity sensor connected to the RS-485 interface
(terminals 1 … 5):
Parameter: Auto or Conductivity (Induct.)
Mode: Other Digital
Functionality: SE680K
Selecting a Second Digital Toroidal Conductivity Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [II]
Selection of a second Memosens toroidal conductivity sensor connected to the MK-MS 095N
measuring module:
Module: MK-MS
Parameter: Conductivity (Induct.)
Mode: Memosens
Functionality: CondI
Selection of a second SE680 digital toroidal conductivity sensor connected to the MK-MS 095N
measuring module:
Module: MK-MS
Parameter: Conductivity (Induct.)
Mode: Other Digital
Adjustable parameters for digital or Memosens sensors for toroidal conductivity
Parameter Setting
[I] [II] Digital/Memosens CondI
Input Filter Pulse Suppression Enable/disable suppression of interference pulses.
Sensor Data
Sensor Data, p. 85
Sensoface Enable/disable display of the Sensoface icon.
With the SE680 sensor:
Sensocheck
Monitoring of primary and secondary coils.
Enable or disable Sensocheck to generate failure or main-
tenance required messages.
With Memosens:
Sensor Monitoring
Details
Option to enter individual limits for monitoring the cell
factor.
Sensocheck: Monitoring of primary and secondary coils.
Enable or disable Sensocheck to generate failure or main-
tenance required messages.
Option to enter individual limits before a message for SIP
counter, CIP counter, and sensor operating time is trig-
gered.