Stratos Multi E401N
84
Adjustable parameters for digital or Memosens sensors for toroidal conductivity
Parameter Setting
[I] [II] Digital/Memosens CondI
Cal Presettings Presetting of calibration mode with corresponding parameters.
➜
Calibration Presettings, p. 86
TC Process Medium Temperature compensation of process medium.
Concentration
➜ Concentration (TAN Option FW-E009), p. 87
TDS Enable/disable TDS function
(“Total Dissolved Solids” that influence conductivity)
USP Enable/disable USP function for monitoring ultrapure water and set the USP limit.
➜
USP Function, p. 87
Messages Enable/disable messages for the separate process variables, or specify individual limits.
➜
Messages, p. 88
Selecting an Analog Toroidal Conductivity Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [II]
Selection of a toroidal conductivity sensor connected to the MK-CONDI035N measuring module:
Module: MK-CONDI
Mode: Analog
Adjustable parameters for analog toroidal conductivity sensors
Parameter Setting
[II] Analog CondI
Input Filter Pulse Suppression Enable/disable suppression of interference pulses.
Sensor Data
➜
Sensor Data, p. 85
Sensor Type Select the used sensor type: SE655, SE656, SE660, ISC40,
ISC40S, 5000TC, other
If you select “Other”, please enter further sensor data.
Sensoface Enable/disable display of Sensoface messages and icons.
Sensocheck Monitoring of primary and secondary coils. Enable or dis-
able Sensocheck to generate failure or maintenance re-
quired messages.
Temperature Detection Select temperature probe, set measuring and calibration
temperature.
Cal Presettings Presetting of calibration mode with corresponding parameters.
➜
Calibration Presettings, p. 86
TC Process Medium Temperature compensation of process medium.
Concentration
➜ Concentration (TAN Option FW-E009), p. 87
TDS Enable/disable TDS function
(“Total Dissolved Solids” that influence conductivity)
USP Enable/disable USP function for monitoring ultrapure water and set the USP limit.
➜
USP Function, p. 87
Messages Enable/disable messages for the separate process variables, or specify individual limits.
➜
Messages, p. 88