Stratos Multi E401N
164
Saturation
1)
0.000...150.0%
Concentration
1)
(dissolved oxygen)
0000...9999µg/l /10.00...20.00mg/l
0000...9999ppb/10.00...20.00ppm
Volume concentration in gas 000.0...9999ppm/1.000 ...50.00vol%
Trace sensor “001” (analog)
Saturation
1)
0.000...150.0%
Concentration
1)
(dissolved oxygen)
000.0...9999µg/l /10.00...20.00mg/l
000.0...9999ppb/10.00...20.00ppm
Volume concentration in gas 000.0...9999ppm/1.000...50.00 vol%
Input correction
Pressure correction 0.000...9999bar/999.9kPa/145.0psi (adjustable)
Manual or external (via current input 0(4)...20mA)
Salinity correction 0.0...45.0g/kg
Calibration and adjustment
Automatic calibration in air-saturated water
Automatic calibration in air
Product calibration, saturation (with offset on SE 740)
Zero correction
Temperature probe adjustment
Calibration ranges
Standard sensor
Zero point ±2nA
Slope 25...130nA (at 25°C/77°F, 1013mbar)
Trace sensor “01”
Zero point ±2nA
Slope 200...550nA (at 25°C/77°F, 1013mbar)
Trace sensor “001”
Zero point ±3nA
Slope 2000...9000nA (at 25°C/77°F, 1013mbar)
Calibration timer
0000...9999h
1)
For temperature range -10 ... 80°C/14 ... 176°F