The system will enter Error state. If the user pressesthe STANDBY button or the
READY button, the system will enter STANDBY state if the triggering condition is no
longer present.
• Monitored Errors
The system will enter Error state. If the triggering condition disappears, the system
will enter Standby state. In all cases, the system will display appropriate messages on
the control panel. The remote control will display the state of the system, but willnot
display error messages.
c. Specific Errors
• Errors of the type indicated will be set under the conditions as indicated in the following
table:
d. Pulse Energy 20% Error Indication.
If the moving average of the pulse energy is greater than 1.2* requested energy or less than
0.8* requested energy, the system will give a distinctive audible beep. The moving average
will be calculated as follows:
Error Error type Condition
Energy High Clearable Energy of any pulse greater than 1.5* requested energy.
Energy Low Clearable Energy of any pulse less than 0.5* requested energy.
Pulse rate high Clearable Interval between two consecutive pulses, as measured by
the independent clock, less than 0.8* requested pulse inter-
val.
Pulse rate low Clearable Interval between two consecutive pulses, as measured by
the independent clock, greater than 1.2* requested pulse
interval.
+5V power supply error Clearable Power supply output voltage differs from nominal by
greater than 10%.
+15V power supply error Clearable Power supply output voltage differs from nominal by
greater than 20%.
-15V power supply error Clearable Power supply output voltage differs from nominal by
greater than 20%.