OPERATIONAL MONITORING:
1. System Monitoring During Standby State
a. In Standby state, the system will monitor the following:
• Footswitch connection
If no footswitch is connected, the system shall display a message and remain in Fault
state.
When a working footswitch is connected to a dual-wavelength system, the system
will display whether it is a single or double-footswitch. This requirement applies only
to dual-wavelength systems.
When a working footswitch is connected and the only error in effect is that there is no
footswitch, the system will go to Standby state automatically.
If the footswitch is defective (if both contacts are in the same state for more than a few
seconds) the system will display a message and go to a Fault state.
• Remote Interlock Connection
If the remote interlock is disconnected, the system will display a message and remain
in a Fault state.
When the remote interlock is reconnected, the system will go to a Standby state auto-
matically.
• Fiber Connection
If the delivery fiber is disconnected, the system will display a message and remain in a
Fault State.
When the delivery fiber is reconnected, the system will go to a Standby state automat-
ically.
• Safety Shutter Position