4. Error Handling.
a. Errors vary in level of severity and the nature of the response required. The VersaPulse PowerSuite
error handler will recognize three classes of errors.
Permanent
The error can be cleared only by power cycling the system.
Clearable
The error can be cleared, the error message will no longer be displayed and the system
will go Standby state when the user presses the READY button or the STANDBY but-
ton, if the tiggering condition is no longer present.
Monitored
The error can be cleared, the error message will no longer be displayed and the system
will go to Standby state when the triggering condition is no longer present.
b. Actions taken
Permanent Errors
The system will enter Error state and remain there. The system will exit Error state
only when the power is cycled andthe triggering condition is no longer present.
Clearable Errors
Dual 80/100JPN Single 803JPN
Dual 30/100JPN Single 30JPN
Dual 20/100JPN Single 20JPN
Dual 201/100 System in non-functional
Dual 201/60 System is non-functional