If a Holmium laser rod fails, the sytem will not attempt to continue to fire the failed rod, but
will continue functioning as if it were a model having a number of Holmium rods actually
available, and maximum pulse rate, pulse energy and output power not greater than those
available originally to the model in use. The system will display a special screen advising the
user to call Lumenis Field Service and informing the user of the maximum available pulse rate
and average power.
If the Nd:YAG rod in a dual-wavelength system fails, the system will continue functioning as
if it were a single-wavelength model having the number of Holmium rods actually available
maximum pulse rate, pulse energy and output power not greater than those available orgin-
ally to the model in use. The system will advise the user as described above.
In case of failure of a single Holmium rod, the system will behave as indicated in the following
table;
In case of failure of the Nd:YAG rod, the system will behave as indicated in the following
table:
Failing Model Change to Model
Dual 80/100 Dual 30/100
Dual 60/100 Dual 30/100
Dual 30/100 Dual 201/100
Dual 201/100 System is non-functional
Dual 20/60 Dual 201/60
Dual 201/60 System is non-functional
Dual 80/100JPN Dual 30/100JPN
Dual 30/100JPN Dual 20/100JPN
Dual 20/100JPN System is non-functional
Failing Model Change to Model
Dual 80/100 Single 803
Dual 60/100 Single 60
Dual 30/100 Single 30
Dual 20/60 Single 20