OTHER COMPONENTS
Fan Motor/Impeller, Pump Motor, Pump Head, Waterflow Switch, Water Temp Sensors, Fil-
ters (DI, particle, air) and General Plumbing:
• Fan Idle Speed Adjustment
• Perform Operational and Safety Checks
Display:
• Load Logo
• Check Autocalibration Values
• Perform Operational and Safety Checks
The following are common faults which may occur during the operation of the laser.
TEMPERATURE
1. The overheating fault is displayed on the LCD screen.
a. Verify that the coolant is at its proper level. (The upper water reservoir should be
between one-half full.
b. Perform the Fan Speed Controller Adjustment in Section 3.3.2.
c. Verify that the water particle filter is clean. If the filter is dirty or discolored
(brown), drain the laser and replace the particle filter and DI cartridge.
d. Check the cleanliness of the air filter on the bottom of the laser. A dirty air filter
will restrict air flow. (This filter should be vacuumed periodically.
e. Verify that there is adequate space at the bottom of the system when it is in use.
(There may be other carts or equipment in the operating theater next to the Ver-
saPulse PowerSuite reducing the air flow to the laser.
f. Verify that the ambient temperature of the operating environment is below 22•C.
e. Verify that water hoses are not bent, kinked, or otherwise damaged.