16. Plot Calibration
Depressing this button displays the Calibration Plots screen.
17. Fiber Centration.
Depressing this button displays the Fiber Centration screen.
18. Draw Main energy plot
Depressing this button displays the Pulse Energy Plot based on the main energy channel.
19. Draw Safety Energy Plot.
Depressing this button displays the Pulse Energy Plot based on the safety energy channel.
20. Return to Service 2 Screen
This button returns the system to Service Screen 2.
21. Plot Calibration Data
The buttons consist of PLOT ALL, PLOT 1, PLOT 2, PLOT 3 and PLOT 4.
Depressing the PLOT ALL button will display graphs of the calibration curves of all laser rods in use.
Depressing each of the other buttons will result in displaying a graph of the calibration curve of the associated
laser rod.
22. Calibration Formula Display
This display will show for each rod, the coefficients of the calibration curve, to 2 decimal places.
23. Calibration Value Table
This displays a table of pump energy or lamp current for the following laser outputs:
Holmium rods - 0.5, 2.0 and 3.5 joules
Nd:YAG - 100W and 25W